Recommended settings for easyPhaco
The Oertli easyPhaco technology is available in three different tip sizes. Incisions of sub–2-mm coaxial micro-phaco up to standard incision sizes of up to 3.2 mm can be used with easyPhaco.
| Minimum | Venturi pump | Peristaltic pump / SPEEP pump | |||
| Vacuum (mmHg) | Vacuum effect (%) | Flow (ml/min) | Vacuum (mmHg) | ||
| easyTip CO-MICS for incisions of 1.6-1.8 mm | Phaco 1 (Grooving) Phaco 2 | 80 250 | 60 60 | 18 20 | 50 250 | 
| easyTip 2.2 mm for incisions of 2.2-2.4 mm | Phaco 1 (Grooving) Phaco 2 | 100 350 | 80 80 | 18 35 | 80 350 | 
| easyTip 2.8 mm for incisions of 2.8-3.2 mm | Phaco 1 (Grooving) Phaco 2 | 100 350 | 80 80 | 18 35 | 80 350 | 
--> Recommended bottle height: 100 cm above the patient eye
--> Recommended additional settings: Pulse mode, 40 Hz, 60% cooling for Phaco 2.
easyPhaco is a development from Oertli R&D in scientific collaboration with Professor Rupert Menapace, Vienna.
Please contact your sales partner to determine your optimal easyPhaco values.